Corneal ring segments decrease corneal curvature but do not affect biomechanics
J Cataract Refract Surg. 2009;35(10):1761-1767.
Submit a Comment Submit a Comment Print  Print Email this page to a friend  E-mail

Intrastromal corneal ring implantation in keratoconic corneas markedly reduced corneal curvature but did not influence corneal biomechanical characteristics, a study showed.

The authors assessed biomechanical and morphological changes in keratoconic corneas implanted with Intacs intrastromal corneal ring segments (Addition Technology).

"To our knowledge, this is the first study that shows that Intacs placement changes the morphological characteristics of the cornea but not the biomechanical viscoelastic response parameters, such as [corneal hysteresis] and [corneal resistance factor]," the authors said.

Story continues below

ADVERTISEMENT

The retrospective study included 18 eyes of 18 patients with a mean age of 31.3 years (range: 13 to 50 years). Patients underwent a complete ocular examination before surgery and 6 months after surgery. Investigators performed corneal topography with the Orbscan II (Bausch & Lomb) and assessed biomechanical factors with the Reichert Ocular Response Analyzer.

Mean preoperative corneal hysteresis was 7.7 mm Hg and mean postoperative corneal hysteresis was 7.4 mm Hg. Mean preoperative corneal resistance factor was 6.6 mm Hg and mean postoperative corneal resistance factor was 6.1 mm Hg. Neither reduction was statistically significant.

"In our study, both IOP and central pachymetry remained stable postoperatively and thus played a role in the lack of a significant change in [corneal hysteresis] and [corneal resistance factor] values," the authors said. "We hypothesize that because of the lamellar structure of the cornea, the [intrastromal corneal ring] segments might reshape the center of the cornea without changing the fundamental biomechanical properties of the corneal tissue, at least in the short-term postoperative period."

However, mean minimum and maximum central keratometry values decreased more than 5 D; the differences were statistically significant (P < .0001).

Further study with a larger patient group and longer follow-up interval is needed to determine the effect of Intacs implantation on corneal hysteresis and corneal resistance factor, the authors said.

The OSN SuperSite is intended for physician use and all comments will be posted at the discretion of the editors. We reserve the right not to post any comments with unsolicited information about medical devices or other products. At no time will the OSN SuperSite be used for medical advice to patients.

There are no comments for this article. Be the first to comment.

Your comment

Name:
Comments:
SEE ALSO
Avaliação da curvatura corneana posterior importante em candidatos à LASIK
OCULAR SURGERY NEWS LATIN AMERICA EDITION December 1, 2009
Cómo evaluar la curvatura posterior de la córnea adquiere importancia en los candidatos a la cirugía LASIK
OCULAR SURGERY NEWS LATIN AMERICA EDITION December 1, 2009
Epithelial ingrowth reduced with femtosecond laser-cut LASIK flaps
Posted on the OSN SuperSite November 20, 2009
Case presentation: Decentered IOL
Case presentation: Decentered IOL

Submitted by: Robert H. Osher, MD (7/9/2009).

Aging population contributes to rising glaucoma burden around the world
Reducing the occurrence of glaucoma poses a challenge around the world, with nearly 8 million people bilaterally blind from the disease.

Web Statistics